test volume: 22m³
working chamber dimensions (W×D×H): 2500×4000×2250mm
overall dimensions (W×D×H): 4400×4480×2650mm
power supply: AC380V ±10%, 50Hz ±1, three-phase four-wire + grounding wire (3/N/PE), grounding resistance < 4Ω
max operating power: approx 65.2KW (subject to final engineering design)
max operating current: approx 113.9A (subject to final engineering design)
cooling method: Water Cooling
temperature range: -40℃ ~ +100℃
temperature uniformity: ≤2℃
temperature fluctuation: ±0.5℃
temperature deviation: ±2℃
heating rate: 3℃/min (average over full range)
cooling rate: 2℃/min (average over full range)
temperature range for humidity: 10℃ ~ 85℃
humidity range: 10%RH ~ 98%RH
humidity uniformity: ±5.0%RH
humidity fluctuation: ±3.0%RH
humidity deviation: +2% ~ -3%RH (>75%RH)
applicable standards: GB/T 2423.1-2008, GB/T 2423.2-2008, GB/T 2423.3-2008, GB/T 2423.4-2008, GB150.3A-2009, GB150.4A-2009, GB150.9A-2009
core functions: 32-bit color screen control, Ethernet & USB data access, APP remote management, over-temperature/smoke/power-off protection, automatic water supply & water shortage alarm, long-term low-temperature defrost-free operation