test capacity: 50L
chamber dimensions (W×D×H): 420×420×320mm
overall dimensions (W×D×H): approx 1170×2600×2000mm (excluding protruding parts)
power supply: AC380V ±10%, 50Hz ±1, 3-phase 4-wire + grounding (3/N/PE), grounding resistance <4Ω
max system power: approx 24 kW (subject to final engineering design)
max system current: approx 40 A (subject to final engineering design)
low temperature shock range: -40℃ ~ -10℃
high temperature shock range: +60℃ ~ +150℃
temperature uniformity: ≤2℃
temperature fluctuation: ±0.5℃
temperature deviation: ±2℃
low temperature ramp range: -20℃ ~ -10℃
high temperature ramp range: 60℃ ~ 100℃
stress speed time: <10 sec (measured at air outlet)
heating temperature range: +50℃ ~ +200℃, heat-up time (ambient → +200℃): approx 40 min
cooling temperature range: -80℃ ~ +70℃, cool-down time (ambient → -80℃): approx 80 min
temperature recovery time: <5 min
temperature transfer time: <10 sec
sample heat dissipation requirement: Recommended heat dissipation of test sample < 300 W, customized solution available for special demands
applicable standards: GB/T 2423.22, MIL STD 202F Method 107G, IEC 60068-2-14 (Test Na), BS 2011, DIN 40046 (Test Na), JESD22-A101-A